Only one device was received for evaluation.No other samples were received and therefore, could not be evaluated.A visual inspection was done which observed the tube set tubing was received with the detached spike found loose in the opened primary packaging.Since the spike was received detached from the tube set tubing, functional testing could not be performed.The reported condition was verified.The cause was not determined; however, the most likely cause was due to inadequate adhesive being applied in the manufacturing process at the spike connection to the tubing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|