• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Injury (2348)
Event Date 01/13/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the perforator did not stop causing a dura tear that had to be repaired.A one hour surgical delay was reported.The perforator clicked in place with the drill and the recommended spring tests were performed between each burr hole.
 
Manufacturer Narrative
The perforator was not returned for evaluation (as per customer, product not available) therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key11274388
MDR Text Key230114781
Report Number3014334038-2021-00017
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ63Z09
Was Device Available for Evaluation? No
Date Manufacturer Received03/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DRILL-NSK PRIMADO 2; DRILL-NSK PRIMADO 2
Patient Age68 YR
-
-