• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 58G; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 PPS LTD ACET SHELL 58G; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Device Problem Noise, Audible (3273)
Patient Problem Pain (1994)
Event Date 08/21/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).D.O.B: unknown month and day in (b)(6).Concomitant medical products: catalog#: 192114, echo por fmrl lat nc 14x150mm, lot#: 133930.Foreign report source: (b)(6).The device will not be returned for analysis as the device remains implanted; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-00334.
 
Event Description
It was reported that approximately 2 years post implantation, the patient was experiencing an increase in pain with movement and frequent noise.Patient states noise is getting worse, and was prescribed physical therapy.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: catalog#: 192114 echo por fmrl lat nc 14x150mm lot#: 133930.Catalog#: 650-1162 delta cer fem hd 32/0mm t1 lot#: 2016120241.Catalog#: 110003629 biolox delta cer lnr 32mm g lot#: 3409704.The event was confirmed with medical records and radiographs received.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 58G
Type of Device
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11274521
MDR Text Key230119962
Report Number0001825034-2021-00335
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000666
Device Lot Number3726509
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age59 YR
Patient Weight104
-
-