MAUDE Adverse Event Report: BIOMET MICROFIXATION SCREW 2.0X8MM X-DRIVE 5PK; PROSTHESIS, MICROFIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION SCREW 2.0X8MM X-DRIVE 5PK; PROSTHESIS, MICROFIXATION Back to Search Results
Catalog Number 45-2008
Device Problem Fracture (1260)
Patient Problems Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Event Date 12/30/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that approximately five (5) weeks ago, the head of the screw fractured off during insertion.No health consequences were reported for the patient as a result of the event.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The investigation is still in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Visual examination of the provided pictures showed that the screw fractured where the screw shaft meets the screw head.Medical records were not provided.Lot identification is necessary for review of device history records, lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW 2.0X8MM X-DRIVE 5PK
Type of Device
PROSTHESIS, MICROFIXATION
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key11280483
MDR Text Key238622397
Report Number0001032347-2021-00058
Device Sequence Number1
Product Code JEY
Combination Product (y/n)N
PMA/PMN Number
K081067
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup,Followup
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number45-2008
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age33 YR
-
-