Model Number SPL-S
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device was evaluated.Inspection of the device confirmed the reported failure.The device transducer receptacle was found damaged and cannot be plugged.Minor scratches and dents on the housing were observed.The reported issue was found to be due to damaged receptacle unit.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
It was reported that the device transducer cannot plug into the generator.The front panel connector was found damaged.The issue occurred during preparation for use.
 
Manufacturer Narrative
This report is being supplemented to inform that upon further review, this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE-SE LITHOTRIPSY SYSTEM
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough PA 01772
MDR Report Key11286026
MDR Text Key230550045
Report Number3011050570-2021-00042
Device Sequence Number1
Product Code FEO
UDI-Device Identifier00821925044197
UDI-Public00821925044197
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-