U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770555
Device Problems Break (1069); Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/23/2020
Event Type  malfunction  
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of cable damage was confirmed.The probe strain relief is cut.The root cause of the failure was identified as use-related damage.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Sales rep reported the probe cable is cut/damaged.01/11/2021- sample evaluation confirmed wire break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
shelly gilbert
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key11287487
MDR Text Key231616979
Report Number3006260740-2021-00250
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098130
UDI-Public(01)00801741098130
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770555
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-