Udi: (b)(4).The certas valve was returned for evaluation: failure analysis - the valve was visually inspected, no defects were noted.The valve passed the test for programming, occlusion, leak, siphon guard.The valve failed the test for reflux, pressure.The valve was dismantled and images were taken at appropriate magnification: biological debris were noted on the flat spring of cam/ball and on the ruby ball.The root cause for the reflux and pressure problems noted during the investigation was due to biological debris on the flat spring of cam/ball and on the ruby ball, the debris was stopping the ruby ball from sitting correctly on the seat of the ruby ball.The possible root cause for the issue reported by the customer "csf did not flow through certas valve" was probably due to biological debris and protein buildup interfering with the valve mechanism.
|