MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON FENESTRATED SCREW SET; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 55750025535
Device Problem Leak/Splash (1354)
Patient Problems Extravasation (1842); Unspecified Tissue Injury (4559)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider via a manufacturer representative regarding a patient with pre-operative diagnosis of vertebra fracture.It was reported that intra-op, the cement leak out into screw head instead of coming out at fenestrated screw holes.The cement was leaked at the lateral part of the vertebra and the surgeons was able to remove it at the end of the surgery with an injury to the patient.Further there were no patient symptoms or complications as a result of this event.There were no injury to patient.Antibiotic and analgesics were prescribed to improve situation.There is no additional device used as a result of the event.There was a delay of 1hr 30 min as a result of the event.
 
Manufacturer Narrative
As it was confirmed there was no injury to the patient during follow up attempt, this should be reported as malfunction instead of serious injury.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON FENESTRATED SCREW SET
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key11292203
MDR Text Key230670897
Report Number1030489-2021-00154
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
PMA/PMN Number
K201362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number55750025535
Device Catalogue Number55750025535
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age62 YR
-
-