• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE EF; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE EF; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Date 01/19/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the tasp bottom became chipped when being washed by sterile processing department.There was no patient involvement.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A visual inspection of the returned item exhibits signs of repeated use (nicked or gouged) the locking feature is fractured.Not all pieces were returned.The dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE EF
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11293172
MDR Text Key230726735
Report Number0001822565-2021-00336
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024245464
UDI-Public(01)00889024245464(11)170205(10)63593884
Combination Product (y/n)N
PMA/PMN Number
K123459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42527600505
Device Lot Number63593884
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2021
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-