Mfr report #: the original mfr report # was 2016493-2020-80002 was submitted with the incorrect mfr registration site number.Although the original mdr referenced the incorrect mfr site the mdr was submitted 21 october 2020 and was on time.(b)(4).Investigation: the customer sent in a texium connector attached to a halyard homeseries c-pump with the complaint of leakage at the texium site.After testing, the leakage could not be duplicated and is likely due to an incompatible connection with bd device.Though texium meets iso specifications, it is advised that the product be used with smartsite bd products.A device history record review for model 10012241-0500, lot number 17126211 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A root cause was not determined as the complaint of leakage could not be replicated.
|
It was reported that 6 texium closed male luer with female cap leaked during use.The following was reported by the initial reporter: "it was reported that there is a leaking issue when using the texium closed male luer with the c-series pump iv line.Describe customer concern: this event has occured 7 times in the last two days and several times earlier this week-last pir submitted.The issue is with our texium, ct # 10012241-0500.Our texium (female end) is being connected to the customer's halyard c-series elastomeric pump.Fixed male swivel luer lock end of the iv line.5fu is being administered and has leaked between the connections.Our texium is being used with all types of homecare sets and applications, ie: buddy pump set but this issue of leaking is only occurring with the c-series pump-iv line.
|