Additional information was added to h3, h4 and h6.H4: the device was manufactured from july 17, 2020 - july 20, 2020.H10: the two (2) devices were received for evaluation.A visual inspection was performed.And it was noted, that there was a crack in the fillport of each sample.Functional leak testing was also performed.And it was noted, that there was a leak observed at the crack in each sample.The reported condition was verified.The cause of the reported condition could not be determined.However, the probable cause of the cracks may have been, due to excess force applied onto the fillport, during filling the device with medication.A batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|