Model Number WCD 4000
Device Problem False Alarm (1013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (gong alarms/ arrhythmia alarms, false asystole events) has been confirmed.Upon investigation the electrode belt failed an ecg fall-off test.The cause for the failure was isolated to an open brown (-5v) wire in the trunk cable.The trunk cable showed signs of physical abuse.The root cause for the open wire was excessive force.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor returned a patient's electrode belt and reported that the patient was experiencing false asystole events, gong alarms, and arrhythmia alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
mary ward
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key11298298
MDR Text Key232082190
Report Number3008642652-2021-01168
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005135
UDI-Public00855778005135
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A1025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2021
Date Manufacturer Received12/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-