MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) Back to Search Results
Model Number DM0010FAA
Device Problem Device Remains Activated (1525)
Patient Problems Perforation (2001); Unspecified Tissue Injury (4559)
Event Date 01/11/2021
Event Type  Injury  
Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was discarded.This dm0010faa easydrill cranial perforator with unknown lot number was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90¿) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90¿).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that when creating a left frontal burr hole for a subthalamic nucleus deep brain stimulation procedure, the perforator did not stop which caused durotomy and cortical injury.They performed hemostasis without problems.It was reported that the patient was tested for a formal neurological exam and the neuro exam was completely normal.It was also reported that the product was discarded by the customer.On follow-up, it was reported that there was a 1-hour procedure delay due to dura repair and the surgery was postponed.It was also reported that the patient had no further issues post-surgery.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
On follow-up, it was reported that the surgery was rescheduled a week later.
 
Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was still in use.D10: serial no# for ad03 is (b)(6) and it is still in use.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR  09980-39
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR   09980-39
Manufacturer Contact
natalia matos
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--39
BR   09980-39
140575722
MDR Report Key11298501
MDR Text Key234691011
Report Number1625507-2021-00025
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K141455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM0010FAA
Device Catalogue NumberDM0010FAA
Device Lot Number723/20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/23/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AD03 (SN (B)(6)); AD03 (UNKNOWN S/N)
Patient Outcome(s) Other;
Patient Age62 YR
Patient SexMale
-
-