This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.As a result of the physical inspection and functional testing of the device olympus has concluded that the damage to the device was likely caused by wear from repeated use over a long period of time.It is also possible that unexpected stress was applied due to mishandling by the user.The device was repaired and functionally tested and returned as an asset to the olympus loaner inventory.Based on the legal manufacturer¿s investigation, the device history record (dhr) was reviewed and there were no problems found during the manufacturing of the device that would cause or contribute to the reported issue.The device met all specifications at the time of shipment.The instructions for use states: ultrasound gastrovideoscope gf-uct180 chapter 3 preparation and inspection 3.2 inspection of the endoscope inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.
|