ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ ASPHERIC UV ABSORBING IOLWITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR
|
Back to Search Results |
|
Model Number ACU0T0 |
Device Problem
Positioning Problem (3009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A health professional reported that during an intraocular lens (iol) implant surgery, the plunger went over the iol.There was patient contact but no patient impact.Additional information has been requested.
|
|
Manufacturer Narrative
|
The used device with the lens was returned.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was observed in the device.The plunger has been retracted to the start point, prior to the lens in the loading area.The lens was completely within the loading area upon return.The trailing haptic was dried into an elevated position.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported event.The used device was evaluated.The sample was not returned in the condition described.The plunger was retracted to the start point, prior to the lens.The lens was still within the loading area with the trailing haptic dried into an elevated position.Based on the condition of the returned sample, it cannot be confirmed if the plunger had been advanced over the lens and then retracted.The manufacturer internal reference number is:(b)(4).
|
|
Search Alerts/Recalls
|
|
|