TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801188 |
Device Problem
Inaccurate Flow Rate (1249)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) verified the reported issue.The fsr found that under normal preventative maintenance (pm) testing the epgs appeared to be functioning properly, but when the epgs was connected to the oxygenator, where some resistance to flow is introduced, the epgs was not compensating for this resistance properly.This resulted in a drop in flow.The epgs was replaced.The unit operated to the manufacturer's specifications.The suspect device was sent back to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the sweep gas flow measurements between the electronic patient gas system (epgs) and the pole mounted flow meter were off.As mitigation, the external flowmeter was used to measure the flow.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the electronic patient gas system (epgs) observed unobstructed flow was acceptable, and when obstructed, the system was able to compensate.Differences in the flow that were observed between the central control monitor (ccm) value and the flowmeter were within the expected tolerance.Per data log analysis, on 15-jan-2021 the gas system was successfully calibrated at 06:22:24.08:25:42 flow was set to 1.5 liters per minute (l/min), 09:25:16 flow was set to 1.7 l/mi,n, 09:40:24 fraction of inspired oxygen (fio2) was set to 0.83 (83%), 09:51:49 flow was set to 2.0 l/min, 09:54:13 flow was set to 2.2 l/min, 10:02:57 flow was set to 0 l/min, 10:54:25 the perfusion screen was exited.There was no indication of a problem in the log.If the flow output did not match the flow reading on the ccm, the flow meter could be faulty.The log did not confirm the complaint.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) could not duplicate the reported issue.The electronic patient gas system (epgs) powered up and established good air and oxygen flow.All values when set at different setpoints were within specification.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Search Alerts/Recalls
|
|
|