• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 10/30/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Investigation completed.This is an initial final report submission.Review of the most recent repair record determined the head, control bar, thickness control lever, width plates 2 and 4 inch were damaged.The width plates, head, control bar, thickness control lever, motor, eccentric shaft, thickness control shaft, fine adjustment cams, planetary carrier, bearings, spring seal, and other parts were replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that upon inspection the device was found in need of repair.At evaluation investigation the device was found to have a damaged width plate; a reportable malfunction.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11307155
MDR Text Key231143729
Report Number0001526350-2021-00146
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number62689593
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-