Model Number GF-UCT180
Device Problem Electrical Power Problem (2925)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The scope was returned to the service center for evaluation.The customer¿s complaint of ¿ nbi button not working¿ was not confirmed.The scope¿s control body was inspected and found the pink rubber chipped.The bending section rubber glue was found chipped and cracked.A leak was noted at scope¿s auxiliary water flow inlet.The elevator channel inlet was found loose.Fluid invasion was noted in the scope¿s control body, scope connector and ultrasound connector.The scope¿s id chip shows 685 uses.The investigation is ongoing and if additional information is received this report will be supplemented accordingly.
 
Event Description
The service center was informed during an unspecified procedure, the scope¿s narrow band imaging (nbi) button was not working.It is unknown if the intended procedure was completed.There was no patient injury.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely some external force was added to the distal end, leading to the peeling of the glue at the tip.In addition, about 8 years and 7 months have passed since the device was manufactured and it may have been affected by aging degradation.The instructions for use have the following statement which can detect the event: "inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11310317
MDR Text Key233830740
Report Number8010047-2021-02501
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356346
UDI-Public04953170356346
Combination Product (y/n)N
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-