OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT260
Device Problems Break (1069); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection of the subject device at the service department of olympus (b)(4) it was found that the foreign material came out from the air/water nozzle and the foreign material clogged in the air/water channel of the subject device.Also it was found that the ultrasound transducer surface was damaged and there was the gap on the adhesive of the distal end.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based on the information from olympus korea (okr), omsc surmied causes of each of the reported phenomenon as follows; the water supply function not working could have occured, as the air/water channel got clogged due to a foreign material entered to the channel or insufficient cleaning.The failure of the ultrasound image could have occurred, as the elements of the ultrasonic probe were missing due to damage to the ultrasonic probe by sharp material or excessive force.The gap on the distal end glue could have occurred due to aging deterioration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11311973
MDR Text Key231690701
Report Number8010047-2021-02510
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberGF-UCT260
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-