Model Number MR850 |
Device Problems
No Audible Alarm (1019); Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint mr850 respiratory heated humidifier is currently en route to fisher & paykel healthcare(f&p) (b)(4) for evaluation.We will provide a follow up report upon completion of our investigation.
|
|
Event Description
|
A distributor in (b)(6) reported that the audible alarm of a mr850 respiratory heated humidifier was faulty.There was no reported patient involvement.
|
|
Event Description
|
A distributor in texas reported that the audible alarm of a mr850 respiratory heated humidifier was faulty.There was no reported patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint mr850 respiratory humidifier was received at our fisher & paykel healthcare (f&p) service centre in california where the device was inspected by a trained f&p personnel.The device was performance tested and the audible alarm was checked.Results: performance testings revealed that the audible alarm of the mr850 respiratory humidifier was faulty.Conclusion: we are unable to determine what may have caused the reported failure.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 heater base.In addition the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
|
|
Search Alerts/Recalls
|