H4: the lot was manufactured from september 3, 2019 - september 4, 2019.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The photograph was visually inspected and showed the tube set was detached from the stressmember which is located at the upper part of the device.The reported condition was verified during initial inspection and due to the nature of the returned sample, no additional testing could be performed.The cause of the condition was not determined; however the most probable cause of the condition is a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|