Catalog Number 2C1008K
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 01/20/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume infusor had a cracked administration tube.This issue was discovered prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from september 3, 2019 - september 4, 2019.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The photograph was visually inspected and showed the tube set was detached from the stressmember which is located at the upper part of the device.The reported condition was verified during initial inspection and due to the nature of the returned sample, no additional testing could be performed.The cause of the condition was not determined; however the most probable cause of the condition is a manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11325474
MDR Text Key232334446
Report Number1416980-2021-00611
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412022512
UDI-Public(01)00085412022512
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C1008K
Device Lot Number19J021
Was Device Available for Evaluation? No
Date Manufacturer Received03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-