Catalog Number A1108 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that based on an inspection report, the swivel lock on the mayfield triad skull clamp (a1108) was very loose.
|
|
Manufacturer Narrative
|
Mayfield triad skull clamp (a1108) was returned for evaluation.The reported complaint was confirmed.Evaluation showed that the lock has rotational and lateral movement and a residue buildup is present.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.Device exceeds its expected life of 7 years (manufactured in 2010).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|