• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Mechanical Problem (1384)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user facility that during the endoscopic ultrasound-guided fine needle aspiration (eus-fna) using the subject device at the user facility, the forceps elevator of the subject device could not be completely moved down to initial position.The user facility completed the procedure with the subject device.The field service engineer checked the subject device on-site and found that the reported phenomenon was duplicated.There was no report of patient injury associated with this event.The service department of olympus (b)(4) checked the subject device and found that the reported phenomenon was not duplicated, and that the forceps elevator of the subject device could move without any problem.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) for evaluation but was returned to olympus thailand (oth).Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based upon the information from oth and similar past cases, there was the possibility that the reported event was attributed to the adhesion of the foreign material to the forceps elevator wire or the forceps elevator wire channel, which might be caused by user's insufficient reprocessing.If additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11325993
MDR Text Key233382399
Report Number8010047-2021-02600
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170341809
UDI-Public04953170341809
Combination Product (y/n)N
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-