Model Number N/A |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Bone Fracture(s) (1870); Pain (1994); Osteopenia/ Osteoporosis (2651)
|
Event Date 02/01/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that patient underwent a procedure 21 days post implantation due to pain and a fractured tibia bone.The fracture was fixed with an implant plate.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).This complaint was confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.X-rays were provided and reviewed by a health care professional.Review found periprosthetic fracture of the medial tibia with slight tibial implant subsidence.Bone quality appears osteopenic.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|