• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE Back to Search Results
Model Number UM-G20-29R-3
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2021
Event Type  malfunction  
Manufacturer Narrative
Olympus medical systems corp.(omsc) could not investigate the device, because the device was not returned to omsc.Since the device was not returned, the exact cause was unknown; however, the following was supposed to be the cause.When the guidewire was slack or looping, the user pulled the device out of the guidewire.And the guidewire port was broken because an external force was applied to the guidewire port.
 
Event Description
During a procedure with the device, the user was unable to pass a guidewire through the guidewire port of the device because the guidewire port of the device was broken.The user replaced the device to another device to complete the procedure.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11330496
MDR Text Key233382024
Report Number8010047-2021-02651
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K001203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUM-G20-29R-3
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-