The reported event was confirmed as manufacturing related.One sample was confirmed to exhibit the reported failure.An amber 3 way foley catheter was returned opened with original packaging.The catheter was attempted to be inflated with 80 ccs of di water using a luer lock syringe.Water began to spray from the drainage eye.Lumen to lumen leakage is the cause of the reported failure.Sample does not meet specification, as it would fail functional leak testing.A potential root cause for this failure could be "operator error".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not required as the reported event is confirmed as manufacturing related.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|