TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801764 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the facility's biomedical engineer, the heart lung machine (hlm) was charged overnight but still would not run on battery.He measured the battery voltages and found them to be 12.3 volts (v) and 8.9 v.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit would not operate on battery.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was not verifiable.Multiple diligence attempts were unsuccessful for part return so an evaluation could not be completed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|