Model Number OER-PRO |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
An olympus engineer helped the customer troubleshoot the problem.The user facility determined that they had an issue with one of the acecide bottles which likely was deformed during improper storage and this was causing the drawer on the device not to close.If additional information is obtained, a supplemental report will be filed.
|
|
Event Description
|
A user facility reported that they were unable to close the acecide drawer on an oer-pro endoscope reprocessor.There was no patient involvement or injuries reported.No additional information has been obtained.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the results of the manufacturer's investigation.As part of the investigation, a review of the device history record (dhr) and a review of the instructions for use (ifu) were conducted.The dhr review did not find any abnormalities or anomalies identified during production.The device met specifications upon release.The root cause of the drawer not closing was a deformed bottle.The bottle was likely deformed during storage.The ifu contains the following statement: - "if any irregularity is detected during an inspection or if the device is clearly malfunctioning, do not use it.Contact olympus for repair.".Olympus will continue to monitor the field performance of this device.
|
|