• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 01/21/2021
Event Type  malfunction  
Manufacturer Narrative
An olympus engineer helped the customer troubleshoot the problem.The user facility determined that they had an issue with one of the acecide bottles which likely was deformed during improper storage and this was causing the drawer on the device not to close.If additional information is obtained, a supplemental report will be filed.
 
Event Description
A user facility reported that they were unable to close the acecide drawer on an oer-pro endoscope reprocessor.There was no patient involvement or injuries reported.No additional information has been obtained.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the manufacturer's investigation.As part of the investigation, a review of the device history record (dhr) and a review of the instructions for use (ifu) were conducted.The dhr review did not find any abnormalities or anomalies identified during production.The device met specifications upon release.The root cause of the drawer not closing was a deformed bottle.The bottle was likely deformed during storage.The ifu contains the following statement: - "if any irregularity is detected during an inspection or if the device is clearly malfunctioning, do not use it.Contact olympus for repair.".Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11340091
MDR Text Key232809372
Report Number8010047-2021-02775
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-