It was reported that the water could not be extracted during the test and hence, the complainant changed a new one.Per follow up via ibc on 09feb2021, there was no information available regarding how the water extracted from foley catheter but no deviation from current hospital procedure.
|
The reported event was inconclusive.No sample was returned for evaluation.The potential root cause for this failure mode could be user related, example: over aspirated, incorrect syringe)/collapse lumen/sac close eye/valve damage.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "to deflate catheter balloon.Gently, insert a syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If the tails, contact an adequately trained professional for assistance as detected by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
It was reported that the water could not be extracted during the test and hence, the complainant changed a new one.Per follow up via ibc on (b)(6) 2021, there was no information available regarding how the water extracted from foley catheter but no deviation from current hospital procedure.
|