• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Incorrect Measurement (1383)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that during surgery the device was cutting too deep.The surgery was delayed 10 minutes and the patient was under anesthesia.No adverse events were reported as a result of the event.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d8, d9, g3, g6, h2, h3, h4, h6, h8, h10.Review of the most recent repair record determined the control bar was not in the correct position, the device was out of calibration, and did not meet the visual inspection; however, the repair was not completed because the customer requested the device be returned unrepaired.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key11348114
MDR Text Key232535879
Report Number0001526350-2021-00144
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number32346500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received04/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-