• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: X7-2T; TRANSDUCER, ULTRASONIC, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

X7-2T; TRANSDUCER, ULTRASONIC, DIAGNOSTIC Back to Search Results
Model Number TRANSDUCER X7-2
Device Problem Mechanical Jam (2983)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event Description
A customer reported an x7-2t model transducer had an articulation issue during use.There was no injury associated with this event.
 
Manufacturer Narrative
Evaluation of the returned transducer confirmed the articulation issue as described by the customer.Destructive testing revealed a broken steering cable and damage to the i-tube.The steering cable had pulled from the solder joint which resulted in the loss of articulation.The evaluation also found the articulation assembly on this transducer exceeded its expected life cycle.The failed transducer was replaced by the philips service engineer at the customer site and no similar issues have been reported post repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X7-2T
Type of Device
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
MDR Report Key11353257
MDR Text Key233474196
Report Number3019216-2021-10022
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00884838061668
UDI-Public00884838061668
Combination Product (y/n)N
PMA/PMN Number
K132304
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTRANSDUCER X7-2
Device Catalogue Number989605414122
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date01/29/2021
Date Manufacturer Received02/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-