The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.The device had not met specifications.Visual evaluation of the returned sample noted one opened (without original packaging), unused bulb syringe.Visual inspection of the sample noted that there was a 2.50 sq mm green silicone-like material within packaging, but outside of the syringe.This does not meet the specification as "product and package (kit) must be free from visible loose or embedded foreign matter greater than an aggregate total of 0.6mm2 or 1/16¿ in length.A potential root cause for this failure mode could be ¿no following to production areas cleaning procedure¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.Labeling review was not performed because labeling could not have prevented the reported failure.Corrections: d, h.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|