MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET II CONSOLE; LAVAGE, JET Back to Search Results
Model Number 66800039
Device Problems Display or Visual Feedback Problem (1184); Defective Component (2292); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that the nurse was setting up the versajet for a surgery and when he turned the unit on he noticed the following: the power display was blank- no numbers showing in the screen.The footswitch connect indicator led was illuminated.System fault indicator was also illuminated.When the pump cartridge was turned from unlocked to locked, the green light wasn't illuminated.Press on footswitch pedal and nothing happened.The versajet was not used for the procedure, and there was no harm to the patient.The procedure was completed using curettes and scalpel with no delay.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device that was intended for use in treatment was not returned for evaluation, with all information provided we have not been able to establish a relationship between the device and the reported event or determine a root cause.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.The associated risk files contain the reported failure/harm or event.However, as no harm has been alleged then additional review is not required.Instructions for use (ifu) contains recommendations and precautionary statements for proper use of product.Factors that are known to contribute to the alleged fault/failure may be a connection issue or component failure.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET II CONSOLE
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11364796
MDR Text Key233035109
Report Number8043484-2021-00379
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00040565124759
UDI-Public00040565124759
Combination Product (y/n)N
PMA/PMN Number
K143115
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66800039
Device Catalogue Number66800039
Was Device Available for Evaluation? No
Date Manufacturer Received06/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-