Model Number 006173P
Device Problem Inflation Problem (1310)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the balloon would not inflate.Stated that the product was inspected prior to insertion so the incident did not reach the patient.
 
Event Description
It was reported that the balloon would not inflate.Stated that the product was inspected prior to insertion so the incident did not reach the patient.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.The product can be used for diagnostic and treatment purposes but was not used on a patient.It was unknown whether the product had caused the reported failure.A potential root cause for this failure could be "blocked inflation lumen".The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage.In case of catheters with a balloon, under sterile conditions, remove the protective sheath and inflate the balloon with 1.5 ml of air or carbon dioxide.Use the inflation syringe included in the package.Completely deflate the balloon after the test.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
023 5F 110CM BIPOLAR BAL ELE
Type of Device
TEMPERATURE PACING ELECTRODE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11365450
MDR Text Key233059406
Report Number1018233-2021-00619
Device Sequence Number1
Product Code LDF
UDI-Device Identifier00801741011207
UDI-Public(01)00801741011207
Combination Product (y/n)N
PMA/PMN Number
K800298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Model Number006173P
Device Catalogue Number006173P
Device Lot NumberGFEW2706
Was Device Available for Evaluation? No
Date Manufacturer Received06/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-