INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿ Back to Search Results
Catalog Number A3059
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Unique device identifier: (b)(4).Device history record: the dhr shows no abnormalities related to the reported failure.The mayfield skull clamp (a3059) was returned for evaluation: the reported complaint was confirmed via inspection of the unit, with the lock having up and down movement and the teeth were grinding when rotated.General maintenance and cleaning is required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
A facility reported that the mayfield composite series skull clamp (a3059 ) was wiggling a lot when placed on a patient.No patient injury or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key11369161
MDR Text Key243907780
Report Number3004608878-2021-00124
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2021
Date Manufacturer Received02/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-