This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the foreign matter could not be completely cleaned and remained because the cleaning at the facility after the previous case was insufficient.In addition, black foreign matter is considered to be the adhesive applied when assembling the equipment.It is presumed that this adhesive was peeled off due to excessive external force applied during reprocessing.The instruction manual for this product describes the reprocessing method around the forceps table, chapter 7 cleaning, disinfection, and sterilization procedures: "brushing around the forceps elevator and instrument channel outlet.Flushing the interior of the forceps elevator." the event can be prevented by performing the reprocess according to the instruction manual.
|