• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 36MM G; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 36MM G; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/02/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical devices: 010000666 ¿ g7 shell ¿ 6709452.Customer has indicated that he product will not be returned to zimmer biomet for the investigation as the product was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a hip procedure, the liner would not fit into the cup.A new liner was opened and able to be used with the same shell.There was no reported harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h4; h6 reported event was unable to be confirmed due to limited information provided by the customer.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 36MM G
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11377795
MDR Text Key233553158
Report Number0001825034-2021-00571
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304524996
UDI-Public(01)00880304524996(17)250129(10)6738224
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000742
Device Lot Number6738224
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age66 YR
Patient Weight122
-
-