One single flothru dpt kit with pressure tubing and merit flush device with yellow actuator was recieved for evaluation.The reported event of leakage issue was not confirmed during evaluation.No leakage was observed from the returned kit during leak test.However, a crack was observed at housing of one sample site during visual examination.Crack was approximately 5 mm in length.Other evaluation was opened to document further investigation at supplier facility of the sample site.This non-conformance involves a pediatric dpt that had a blood leakage.If this non-conformance were to occur during use, it has the potential to result in minor blood/fluid loss.This device is used for monitoring pediatric arterial blood pressure and blood sampling.Since these devices are indicated for pediatric patients whose blood volume is much smaller than an adult, they cannot tolerate blood loss as easily as adults.A leak in the system may cause a dampened waveform to be displayed on the patient monitor, thereby alerting the clinician to start the troubleshooting process.This non-conformance may cause a delay in the procedure while the product is exchanged.Lot number was not provided, therefore review of the manufacturing records could not be completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|