DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit and found that the pressure was very slow to calibrate and was unable to adjust pressure above 365.The fse troubleshooted the unit and identified that the scroll compressor was defective.As such, the fse resolved the issue by replacing the scroll compressor and filters.The fse then successfully performed drive calibration, as well as pm with full calibration, all functional and safety checks to meet factory specifications.Moreover, as part of pm protocol, the fse replaced the expired main batteries and critical alarm battery.The iabp was then released to the customer and cleared for clinical service.(b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the pressure regulator calibration failed.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the pressure regulator calibration failed.There was no patient involvement, and no adverse event reported.
|
|