Model Number N/A |
Device Problems
Failure to Shut Off (2939); Intermittent Loss of Power (4016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
It was reported that during surgery the electric dermatome worked lazy, stopped running and sometimes was difficult to shut off.No harm or injury to the patient.No additional graft was taken.An alternate device was used to complete the procedure.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the motor's speed was not stable and showed signs of oxidation.The motor was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|