Catalog Number 828804PL
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Dhr - lot 4837610 conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected; the needle guard was raised.The valve passed the test for programming, leak, reflux, siphon guard and pressure.The valve was flushed; no occlusion was noted but the flow was difficult.The needle chamber was dismantled; the needle guard was bent, and glue traces were noted on the needle guard and the silicone housing base.The root cause for the problem reported by the customer is due to the raised needle guard this is probably due to wrong handling as noted in the ifu : do not fold or bend the valve, folding or bending might cause rupture of the silicone housing, needle guard disc dislodgement or occlusion of the fluid pathway.
 
Event Description
2 of 2 reports.Other mfg report number: 3013886523-2021-00096.A surgeon reported an out of box failure of the certas valve.Issue detected before implantation.A minimal surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAS PLUS INLINE WITH SIPHON
Type of Device
CERTAS PLUS W/ SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key11388917
MDR Text Key234254992
Report Number3013886523-2021-00095
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K143111
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number828804PL
Device Lot Number4837610
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2021
Date Manufacturer Received01/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-