One flothru dpt with stopcocks and pressure tubing were returned for examination.The reported event of pressure measurement issue was confirmed.The dpt did not zero or sense pressure on a pressure monitor.Electrical testing showed that that the input impedance met specification, but the output circuit had an open condition.No visible damage was found at the dpt cable connector, sensor chip or solder joints of the dpt.Continuity testing confirmed that leadwires were continuous between the dpt cable connector and the cable near the dpt.The open condition was located between the #3 solder joint and the cable.Electrical testing from the solder joints showed that both input and output impedance were within specification.The above electrical evaluation suggested that the issue was within the dpt cable.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patients clinical manifestations.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|