Model Number N/A |
Device Problem
Material Frayed (1262)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).The investigation is still in progress.Once the investigation is complete a follow up mdr will be submitted.
|
|
Event Description
|
It was reported that the device is stripped.Wires were exposed.There was no harm or delay.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the plug harness assembly was damaged at the connection.The plug harness assembly was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|
Event Description
|
There is no additional information.
|
|