• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE ADULT-PED DRY/ WET LF; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028442
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history record of batch number 74f2000964 that belong to catalog number a-6000-08lf (pe adult-ped dry/ wet lf) has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
There was an abnormal bubbling (not due to patient) that happened with 5 pleur-evac with 5 different patients.There is a leak at the level between the removable drain tube and the plug in the pleur-evac chamber.We figured it out by clamping the drain and there was still bubbling while clamped.We solved the leak issue by doing multiple connections-disconnections-reconnections of this tubing in the pleur-evac until the abnormal bubbling stopped.All current condition of the 5 patients are stable.There is an increase of the time spent in the hospital due to the abnormal bubbling on the drain circuit.All the 5 pleur-evac are still used in the patients in situ.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE ADULT-PED DRY/ WET LF
Type of Device
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key11399565
MDR Text Key234785805
Report Number3004365956-2021-00074
Device Sequence Number1
Product Code KDQ
UDI-Device Identifier34026704631231
UDI-Public34026704631231
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Risk Manager
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/11/2023
Device Model NumberIPN028442
Device Catalogue NumberA-6000-08LF
Device Lot Number74F2000964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-