Model Number SPL-T
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device inspection of the returned transducer found that the suction outlet was damaged.No additional failures were noted.The device is non repairable.The investigation is ongoing; therefore, the root cause of the reported failure cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
It was reported that during an asset return inspection the device was found with damaged suction outlet.There was no patient involvement on this event reported.No user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.All records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Based on device evaluation results, damage to the device's suction outlet was most likely due to user mishandling.The device ifu advises :this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough PA 01772
MDR Report Key11400248
MDR Text Key244399611
Report Number3011050570-2021-00047
Device Sequence Number1
Product Code FEO
UDI-Device Identifier00821925043831
UDI-Public00821925043831
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-T
Device Catalogue NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-