• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADULT INSPIRATORY HEATED BREATHING CIRCUIT; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADULT INSPIRATORY HEATED BREATHING CIRCUIT; BTT Back to Search Results
Model Number RT212
Device Problems Leak/Splash (1354); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The rt212 adult inspiratory heated breathing circuit is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.The complaint (b)(4) adult inspiratory heated breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A distributor in japan reported on behalf of a healthcare facility that a rt212 adult inspiratory heated breathing circuit failed the ventilator leak test before use.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The rt212 adult inspiratory heated breathing circuit is not sold in the usa but it is similar to a product which is sold in the usa.The 510(k) for that product is k983112.Method: the complaint rt212 adult inspiratory heated breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand for evaluation where it was pressure tested.Results: pressure testing revealed that the circuit was out of specification.The water bath test revealed that the leak was through the connection between the lid and bowl of the water trap.Conclusion: we are unable to determine the cause of the reported event.The water trap of the breathing circuit consists of two parts, a lid and a bowl, which can be separated to allow the caregiver to empty the water trap.All breathing circuits are pressure tested for leaks during production and those that fail are rejected.This suggests that any leak must have developed after the breathing circuit was released for distribution, during transport, storage or use, possibly by distortion of the water trap when the bowl was connected.The user instructions that accompany the rt212 adult inspiratory heated breathing circuit state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
 
Event Description
A distributor in japan reported on behalf of a healthcare facility that a rt212 adult inspiratory heated breathing circuit failed the ventilator leak test before use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Type of Device
BTT
MDR Report Key11402273
MDR Text Key242052312
Report Number9611451-2021-00235
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012430359
UDI-Public(01)09420012430359(10)2101197755(11)200713
Combination Product (y/n)N
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT212
Device Catalogue NumberRT212
Device Lot Number2101197755
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2021
Date Manufacturer Received05/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-