One flow thru dpt sensor with pressure tubing was returned for evaluation.The reported event of inaccurate values was not able to be confirmed, as the monitor did not recognize the dpt sensor and no initial pressure value was shown on the pressure monitor.Dpt did not zero nor sense pressure on the pressure monitor.Electrical testing showed that input impedance and output circuits were open conditions.No visible damage was found at dpt cable connector, sensor chip or solder joints of dpt.Continuity testing confirmed that leadwires were continuous between dpt cable connector and cable near dpt, but open condition between #3 and 4 solder joints and cable near dpt respectively.Input and output circuits on sensor chip were continuous.Above electrical evaluation suggested that the issue was within dpt cable.X-ray evaluation showed #4 leadwire of input circuit and #3 leadwire of output circuit were completely broken inside dpt cable within potting area of dpt housing and created open condition.Since the part was manufactured at the supplier facility, the scar-001549 was generated to deeply investigate the dpt cable / broken leadwire condition at the supplier facility.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patients clinical manifestations.It is not known if user or procedural factors may have contributed to the stated event.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|