• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC. THERMOCOOL SMARTTOUCH SF; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC. THERMOCOOL SMARTTOUCH SF; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number D134805
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Output Problem (3005)
Patient Problem Insufficient Information (4580)
Event Date 02/10/2021
Event Type  malfunction  
Event Description
Incident description: ablation catheter stopped working mid case: high impedance reading and noise on distal electrode.Replaced cable, problem not resolved; restarted machine, problem persisted; replaced ablation catheter with different one and problem resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL SMARTTOUCH SF
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC.
15715 arrow hwy.
irwindale CA 91706
MDR Report Key11411665
MDR Text Key234670979
Report Number11411665
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public(01)10846835010183(17)211214(10)30471000M
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD134805
Device Catalogue NumberD134805
Device Lot Number30471000M
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/02/2021
Event Location Hospital
Date Report to Manufacturer03/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age28105 DA
Patient Weight65
-
-