• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP-I
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2021
Event Type  malfunction  
Manufacturer Narrative
A rattling noise inside the cardiohelp unit was reported.A getinge field service technician was sent for investigation on 2021-02-23, he could confirm that the cause for the rattling noise were loose screws.2 screws and their lock washers of the power supply were reattached.After repair all tests were passed successfully.Based on the results the reported failure "rattling noise" could be confirmed.In consultation with getinge r&d and technical services it was determined that the screws are not properly fixated and became looser over the long time of use (unit manufactured in 2013).For this issue the engineering change request ecr 21030104 was initiated on 2021-03-01 to prevent loosening of the screws by applying loctite.The occurrence rate was calculated for the reported failure and it was determined that this is not a systemic issue.Therefore, no remedial action is required the occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
The customer reported a rattling noise inside of a cardiohelp unit.The unit was investigated by a getinge service technician on 2021-02-23 and screws rattling inside the unit were detected.2 screws and 2 lock washers came off from the power supply slide-bar bracket.The screws were reattached.Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key11434709
MDR Text Key239814459
Report Number8010762-2021-00146
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP-I
Device Catalogue Number701048012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-