• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC CHLORAPREP ONE-STEP; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC CHLORAPREP ONE-STEP; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 930500NSB
Device Problems Premature Activation (1484); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4) initial mdr.A follow up will be submitted if additional information becomes available.(b)(4).
 
Event Description
It was reported chloraprep applicator had no liquid.
 
Manufacturer Narrative
Samples were received by our quality team for evaluation.Based on the samples, the incident of pre-activation could be verified.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.Based on the investigation, the most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handing.Ampoules are made from onion tubing skin borosilicate type i glass, which are design to break when pressure is applied to activate applicator at a relatively low break force.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator, breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see narrative below.
 
Event Description
It was reported chloraprep applicator had no liquid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE-STEP
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION, INC
75 n. fairview drive
vernon hills IL 60061
MDR Report Key11435108
MDR Text Key267933882
Report Number3004932373-2021-00097
Device Sequence Number1
Product Code KXG
Combination Product (y/n)Y
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2023
Device Catalogue Number930500NSB
Device Lot Number0189104
Date Manufacturer Received06/28/2021
Patient Sequence Number1
Patient Outcome(s) Other;
-
-